Campbell CS655 Manuale Utente Pagina 18

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 34
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 17
Model HFP01SC Self-Calibrating Soil Heat Flux Plate
;Measure voltage across current shunt resistor (10 ohm 1% 0.25 W 50
;ppm/deg C) during calibration. This measurement is used to
;compute power.
;
8: Volt (SE) (P1)
1: 1 Reps
2: 25 2500 mV 60 Hz Rejection Range
3: 2 SE Channel
4: 7 Loc [ V_Rf ]
5: .001 Mult
6: 0 Offset
9: Z=X (P31)
1: 2 X Loc [ shf_mV ]
2: 5 Z Loc [ mV_180 ]
;Turn off the soil heat flux plate heater.
;
10: Do (P86)
1: 58 Set Port 8 Low
11: End (P95)
;Stop filtering data.
;
12: If time is (P92)
1: 39 Minutes (Seconds --) into a
2: 180 Interval (same units as above)
3: 30 Then Do
13: Do (P86)
1: 28 Set Flag 8 Low
;Compute in-situ calibration.
;
14: Z=X (P31)
1: 2 X Loc [ shf_mV ]
2: 6 Z Loc [ mV_end ]
15: Z=X*Y (P36)
1: 7 X Loc [ V_Rf ]
2: 7 Y Loc [ V_Rf ]
3: 3 Z Loc [ cal ]
16: Z=X*F (P37)
1: 3 X Loc [ cal ]
2: 128.7 F
3: 3 Z Loc [ cal ]
17: Z=X+Y (P33)
1: 4 X Loc [ mV_0 ]
2: 6 Y Loc [ mV_end ]
3: 9 Z Loc [ work ]
18: Z=X*F (P37)
1: 9 X Loc [ work ]
2: .5 F
3: 9 Z Loc [ work ]
12
Vedere la pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 33 34

Commenti su questo manuale

Nessun commento