Campbell HMP45C Guida Utente

Navigare online o scaricare Guida Utente per Misurazione, test & controllo Campbell HMP45C. Campbell HMP45C User guide Manuale Utente

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 0
HMP45C
Temperature and
Relative Humidity Probe
User Guide
Issued 27.5.09
Copyright
©
1990-2009 Campbell Scientific Inc.
Printed under Licence by Campbell Scientific Ltd.
CSL 373
Vedere la pagina 0
1 2 3 4 5 6 ... 27 28

Sommario

Pagina 1 - Relative Humidity Probe

HMP45C Temperature and Relative Humidity Probe User Guide Issued 27.5.09 Copyright © 1990-2009 Campbel

Pagina 2

HMP45C Temperature and RH Probe 2 2.1 Temperature Sensor Sensor: 1000 Ω PRT, IEC 751 1/3 Class B Temperature Measurement Range: -40°C to +60°C Temp

Pagina 3 - Guarantee

User Guide 3 Do not overtighten the nut that holds in the probe as the probe could be damaged. Figure 1. HMP45C and 41003-5 Radiation Shield on a T

Pagina 4

HMP45C Temperature and RH Probe 4 5. Continue to carefully insert the probe a further 10mm, until the shoulder of the larger section is approximately

Pagina 5 - PLEASE READ FIRST

User Guide 5 1. Do not push the probe so far into the radiation shield that the sensor tip touches the top of the shield. 2. Ensure that your last

Pagina 6

HMP45C Temperature and RH Probe 6 Table 1. Connections for Single-Ended Measurements Sensor Connection Colour CR10X CR1000, CR3000, CR800,

Pagina 7 - Contents

User Guide 7 Table 3. Power Connections using PSW12 Peripherals HMP45C PSW12 Peripheral Datalogger Descrip

Pagina 8

HMP45C Temperature and RH Probe 8 Table 6. Wiring for Single-ended Measurement Examples Description Colour CR1000 CR10(X) Temperature White SE 2 (

Pagina 9 - Humidity Probe

User Guide 9 CR10(X) Program using Single-Ended Measurement Instructions Using a PSW12 or SW12V on Datalogger ;Turn the HMP45C on. ; 01: Do (P86) 1:

Pagina 10 - Error ( C)

HMP45C Temperature and RH Probe 10 6. Long Lead Lengths This section describes the error associated with measuring the HMP45C with a single-ended mea

Pagina 11

User Guide 11 CR1000 Program using Differential Measurement Instructions Using SW12 on Datalogger 'CR1000 program to measure HMP45C with differen

Pagina 13 - 4. Wiring

HMP45C Temperature and RH Probe 12 ;Measure the HMP45C temperature. ; 03: Volt (Diff) (P2) 1: 1 Reps 2: 5 2500 mV Slow Range ;CR510, CR500 (

Pagina 14

User Guide 13 in the air has not changed. Thus, the amount of water vapour in the air, relative to saturation, has decreased. Because of the inverse

Pagina 15 - 5. Example Programs

HMP45C Temperature and RH Probe 14 CR10(X) Program that Computes Vapour Pressure and Saturation Vapour Pressure ;Turn the HMP45C on. ; 01: Do (P86)

Pagina 16

User Guide 15 ;Turn the HMP45C off. ; 05: Do (P86) 1: 51 Set Port 1 Low ;Compute the saturation vapour pressure. ;The temperature must be in d

Pagina 17

HMP45C Temperature and RH Probe 16 9. Troubleshooting Symptom: -9999, NAN, -40 deg C, or 0% relative humidity 1. Check that the sensor is wired to t

Pagina 18 - 6. Long Lead Lengths

Appendix A. Wiring for Older HMP45C Probes Description Colour Temperature Signal Yellow Relative Humidity Signal Blue Signal Reference Purple Po

Pagina 21

CAMPBELL SCIENTIFIC COMPANIES Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com info@campbellsci

Pagina 22

Guarantee This equipment is guaranteed against defects in materials and workmanship. This guarantee applies for twelve months from date of delivery.

Pagina 24 - 9. Troubleshooting

PLEASE READ FIRST About this manual Please note that this manual was originally produced by Campbell Scientific Inc. primarily for the North American

Pagina 26

Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to speci

Pagina 27

Tables 1. Connections for Single-Ended Measurements....6 2. Connections for Differential Measurements...

Pagina 28

1 HMP45C Temperature and Relative Humidity Probe 1. General Description The HMP45C Temperature and Relative Humidity probe contains a Platinum Resist

Commenti su questo manuale