HMP45C Temperature and RH Probe
2
2.1 Temperature Sensor
Sensor: 1000 Ω PRT, IEC 751 1/3 Class B
Temperature Measurement Range: -40°C to +60°C
Temperature Output Signal range: 0.008 to 1.0 V
Temperature Accuracy:
0.4
0.2
0.0
-0.2
-0.4
-40 -20
Temperature ( C)
o
Error ( C)
o
0 20 40 60
2.2 Relative Humidity Sensor
Sensor: HUMICAP
®
180
Relative Humidity Measurement Range: 0 to 100% non-condensing
RH Output Signal Range: 0.008 to 1 VDC
Accuracy at 20°C
±2% RH (0 to 90% Relative Humidity)
±3% RH (90 to 100% Relative Humidity)
Temperature Dependence of Relative Humidity Measurement: ±0.05% RH/°C
Typical Long Term Stability: Better than 1% RH per year
Response Time (at 20°C, 90% response): 15 seconds with membrane filter
3. Installation
The HMP45C must be housed inside a radiation shield when used in the field.
The MET21, URS1 (Figure 1) or 41003-5 Radiation Shield (Figure 1 and 2)
mounts to a tripod mast, UT930 ATW tower leg.
A standard lead length of 3 m allows the HMP45C to be mounted at a 2 or 3 m
height on a tripod mast or tower leg.
The black outer jacket of the cable is Santoprene
®
rubber. This
compound was chosen for its resistance to temperature extremes,
moisture, and UV degradation. However, this jacket will support
combustion in air. It is rated as slow burning when tested
according to U.L. 94 H.B. and will pass FMVSS302. Local fire
codes may preclude its use inside buildings.
With the exception of the URS1 (see below) the probe should be installed so the
tip of the probe is positioned in the middle of the shield, relative to top and
bottom plates of the shield.
NOTE