HMP45C Temperature and RH Probe
6
Table 1. Connections for Single-Ended Measurements
Sensor
Connection
Colour
CR10X
CR1000,
CR3000,
CR800,
CR5000,
CR23X
CR10, CR510,
CR500
21X, CR7
Temperature
Signal
White
Single-Ended
Input
Single-Ended
Input
Single-Ended
Input
Single-Ended
Input
Relative Humidity
Signal
Green Single-Ended
Input
Single-Ended
Input
Single-Ended
Input
Single-Ended
Input
Signal Reference Blue AG AG
Power Ground Black AG AG
Shield Shield G G
Power
Continuous/Switched
Red SW12V SW12 12V/PSW12* 12V/PSW12*
CR10X Power Control
if using SW12V
channel on datalogger
Jumper from
SW12V Control to
Control Port
*On these dataloggers switched power is only available with the PSW12 peripheral.
Table 2. Connections for Differential Measurements
Sensor
Connection
Colour
CR10X
CR1000,
CR3000,
CR800,
CR5000,
CR23X
CR10, CR510,
CR500
21X, CR7
Temperature
Signal
White Differential
Input
H
Differential
Input
H
Differential
Input
H
Differential
Input
H
Temperature Signal
Reference
Jumper to
Blue
Differential
Input – L
Differential
Input – L
Differential
Input – L
Differential
Input – L
Relative Humidity Signal Green Differential
Input – H
Differential
Input – H
Differential
Input – H
Differential
Input – H
Signal Reference Blue Differential
Input – L
Differential
Input – L
Differential
Input – L
Differential
Input – L
Power Ground Black G G G
Shield Shield G G
Power
Continuous/Switched
Red 12V/SW12V 12V/SW12 12V/PSW12* 12V/PSW12*
CR10X Power Control
if using SW12V
channel on datalogger
Jumper from
SW12V Control
to Control Port
*On these dataloggers switched power is only available with the PSW12 peripheral.