QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES
TITLE
SERVICING AND CALIBRATION OF OPTICAL MONITORING
DATALOGGERS
TYPE
STANDARD OPERATING PROCEDURE
NUMBER
4250
DATE
MARCH 1994
AUTHORIZATIONS
TITLE NAME SIGNATURE
ORIGINATOR James H. Wagner
PROJECT MANAGER James H. Wagner
PROGRAM MANAGER David L. Dietrich
QA MANAGER Gloria S. Mercer
OTHER
REVISION HISTORY
REVISION
NO.
CHANGE
DESCRIPTION
DATE AUTHORIZATIONS
1.0 Add responsibilities and equipment. May 1996
Vedere la pagina 0
1 2 3 4 5 6 ... 55 56

Sommario

Pagina 1

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF OPTICAL MONITORINGDATALOGGERSTYPESTANDARD OPERATING PROCEDUREN

Pagina 2 - Page i of i

Number 4250-2000Revision 1.0Date MAY 1996Page 1 of 71.0 PURPOSE AND APPLICABILITYThis technical instruction (TI) describes procedures for servicing an

Pagina 3 - Page 1 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 2 of 7• Ensure that all datalogger servicing is documented and archived in accordance with theprocedures

Pagina 4 - Page 2 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 3 of 7• Replacement components as required• Medium screwdrivers (flat-blade and Phillips-head)• Battery

Pagina 5 - Page 3 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 4 of 7CAMPBELL 21X DATALOGGERSERVICING DOCUMENTATION FORMDate: ___________________________Datalogger S/N

Pagina 6 - Page 4 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 5 of 7After a few seconds delay, the memory check results will bedisplayed. If all memory is installed

Pagina 7 - Page 5 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 6 of 7Documentation Form. The correct reading for each port is shown,along with the manufacturers'

Pagina 8

Number 4250-2000Revision 1.0Date MAY 1996Page 7 of 7RECHARGE If the battery voltage is less than 11.76 volts, connect theBATTERY datalogger to the bat

Pagina 9 - Page i of i

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF THE HANDAR 540A/570A DCPTYPETECHNICAL INSTRUCTIONNUMBER4250-20

Pagina 10 - Page 1 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page i of iiTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONSIBILITIES 12.1 Project Man

Pagina 11 - Page 2 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page ii of iiLIST OF FIGURES (CONTINUED)Figure Page4-2 Front Panel Configurations - Handar 540A/570A DCP 54-

Pagina 12 - Page 3 of 7

Number 4250 Revision 1.0 Date MAY 1996 Page i of iTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONS

Pagina 13 - SERVICING DOCUMENTATION FORM

Number 4250-2010Revision 0.1Date MAY 1996Page 1 of 371.0 PURPOSE AND APPLICABILITYThis technical instruction (TI) describes procedures for servicing a

Pagina 14 - Page 5 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 2 of 37• Coordinate with Handar for return, servicing, and recalibration of 540A/570A DCPsthat fail to o

Pagina 15 - Page 6 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 3 of 37• Electronic contacts cleaning fluid• Medium screwdrivers (flat-blade and Phillips-head)4.0 METHO

Pagina 16 - Page 7 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 4 of 37POST-FIELD INSPECTION CHECKLISTHANDAR 540A/570A DCPDCP S/N: ______________________________Site:

Pagina 17

Number 4250-2010Revision 0.1Date MAY 1996Page 5 of 37Figure 4-2. Front Panel Configurations - Handar 540A/570A DCP.

Pagina 18 - Page i of ii

Number 4250-2010Revision 0.1Date MAY 1996Page 6 of 374.1.2 Physical Inspection - ExternalINSPECT Inspect the outside of the DCP case thoroughly. Look

Pagina 19 - Page ii of ii

Number 4250-2010Revision 0.1Date MAY 1996Page 7 of 37Figure 4-3. DCP Component Diagram.

Pagina 20 - Page 1 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 8 of 37SWITCH When the "TERM" setup screen (see Figure 4-4) is displayed onBAUD RATE the compu

Pagina 21 - Page 2 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 9 of 37Figure 4-4. TERM Setup Screen.

Pagina 22 - Page 3 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 10 of 37Figure 4-5. "TERM" DCP Programming Screen (Program in DCP Memory).

Pagina 23 - HANDAR 540A/570A DCP

Number 4250 Revision 1.0 Date MAY 1996 Page 1 of 51.0 PURPOSE AND APPLICABILITYThis standard operating procedure (SOP) ou

Pagina 24 - Page 5 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 11 of 37Figure 4-6. TERM DCP Programming Screen (No Program in DCP Memory).

Pagina 25 - Page 6 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 12 of 37SWITCH Switch the DCP from "RUN" mode to "PROGRAM" mode byMODES entering ?.T

Pagina 26 - Page 7 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 13 of 37COMPARE The DCP data channel 04 reading (Rotronics AT output signal)READINGS must be adjusted by

Pagina 27 - Page 8 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 14 of 37ROUTINE SERVICING CHECKLISTHANDAR 540A/570A DCPDCP S/N: __________________________Date: ________

Pagina 28 - Page 9 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 15 of 37VERIFY Verify that the GOES primary channel select switches are set toGOES “900” and the seconda

Pagina 29 - Page 10 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 16 of 37CLEAN INTERIOR Clean the inside of the DCP with compressed air.OF DCPMODIFY 570A If this is a Ha

Pagina 30 - Page 11 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 17 of 37Figure 4-8. Handar 570A ADC Board - Component Locations.

Pagina 31 - Page 12 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 18 of 37Figure 4-9. Handar 540A Met Board - Component Locations.

Pagina 32 - Page 13 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 19 of 374.3 DCP PROGRAMMINGThe Handar 540A/570A DCP operational configuration is established through a u

Pagina 33 - ROUTINE SERVICING CHECKLIST

Number 4250-2010Revision 0.1Date MAY 1996Page 20 of 37Table 4-1DCP ID AssignmentsIMPROVE Transmissometer NetworkDCP-ID CHAN LOCATION TIME * RATE **FA4

Pagina 34 - Page 15 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 2 of 52.2 INSTRUMENT TECHNICIANThe instrument technician shall:• Perform all

Pagina 35 - Page 16 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 21 of 37EXECUTE Execute the Handar DCP communications program "TERM" fromPROGRAM the PC comput

Pagina 36 - Page 17 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 22 of 37• $ Forced Scan• # Forced Transmit (GOES Radio)Enter I to edit the station ID. The download pro

Pagina 37 - Page 18 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 23 of 37Enter M to select DCP channel 01. The display message will be "PCHANNEL 01."Enter N to

Pagina 38 - Page 19 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 24 of 37LABORATORY PERFORMANCE TESTINGHANDAR 540A/570A DCPDCP S/N: __________________________Date: _____

Pagina 39 - Page 20 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 25 of 374.4.3 Run Mode Timing ChecksVERIFY Verify that the GOES primary channel select switches are set

Pagina 40 - Page 21 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 26 of 37Set the power element of the wattmeter for the forward direction.Monitor the wattmeter reading a

Pagina 41 - Page 22 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 27 of 37APPENDIX AHANDAR 540A DCP CONFIGURATION PROGRAM - 540ROT.DCP

Pagina 42 - Page 23 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 28 of 37P ID FA43D0BEP STATION TIME 15:47:38P YEAR (XX) 94P DCP JULIAN DATE 055P GOES PRI XMT

Pagina 43

Number 4250-2010Revision 0.1Date MAY 1996Page 29 of 37P CHANNEL NO. 03P03 SENSOR TYPE 10P03 CARD SLOT # 06

Pagina 44 - Page 25 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 30 of 37P CHANNEL NO. 06P06 SENSOR TYPE 10P06 CARD SLOT # 06

Pagina 45 - Page 26 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 3 of 5• Reference thermometer (°C)• Replacement components as required• Batte

Pagina 46 - APPENDIX A

Number 4250-2010Revision 0.1Date MAY 1996Page 31 of 37P08 LOW DIFF LIMIT NO LIMITP CHANNEL NO. 09P09 SENSOR TYPE 10

Pagina 47 - Page 28 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 32 of 37APPENDIX BHANDAR 570A DCP CONFIGURATION PROGRAM - 570TROT.DCP

Pagina 48 - Page 29 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 33 of 37P ID FA43D0BEP STATION TIME 22:46:25P YEAR (XX) 94

Pagina 49 - Page 30 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 34 of 37P02 MEAS INTERVAL 01:00:00P02 START OF MEAS 17:30:00P02 LEVEL 1 MEAS TYPE 001

Pagina 50 - Page 31 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 35 of 37P05 SENSOR TYPE 10P05 SENSOR NAME TAG 10P05 CARD SLOT # 01

Pagina 51 - APPENDIX B

Number 4250-2010Revision 0.1Date MAY 1996Page 36 of 37P07 ZERO SCALE 000P07 MEAS INTERVAL 01:00:00P07 START OF MEAS 17:30:00

Pagina 52 - Page 33 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 37 of 37P CHANNEL NO. 10P10 SENSOR TYPE 12P10 SENSOR NAME TAG 12

Pagina 53 - Page 34 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 4 of 5• Frequency counter• Standard electronics laboratory small tools• Solte

Pagina 54 - Page 35 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 5 of 5• Routine laboratory servicing• DCP programming• Pre-field performance

Pagina 55 - Page 36 of 37

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF CAMPBELL 21X DATALOGGERSTYPETECHNICAL INSTRUCTIONNUMBER4250-20

Pagina 56 - Page 37 of 37

Number 4250-2000Revision 1.0Date MAY 1996Page i of iTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONSIBILITIES 12.1 Project Mana