QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF OPTICAL MONITORINGDATALOGGERSTYPESTANDARD OPERATING PROCEDUREN
Number 4250-2000Revision 1.0Date MAY 1996Page 1 of 71.0 PURPOSE AND APPLICABILITYThis technical instruction (TI) describes procedures for servicing an
Number 4250-2000Revision 1.0Date MAY 1996Page 2 of 7• Ensure that all datalogger servicing is documented and archived in accordance with theprocedures
Number 4250-2000Revision 1.0Date MAY 1996Page 3 of 7• Replacement components as required• Medium screwdrivers (flat-blade and Phillips-head)• Battery
Number 4250-2000Revision 1.0Date MAY 1996Page 4 of 7CAMPBELL 21X DATALOGGERSERVICING DOCUMENTATION FORMDate: ___________________________Datalogger S/N
Number 4250-2000Revision 1.0Date MAY 1996Page 5 of 7After a few seconds delay, the memory check results will bedisplayed. If all memory is installed
Number 4250-2000Revision 1.0Date MAY 1996Page 6 of 7Documentation Form. The correct reading for each port is shown,along with the manufacturers'
Number 4250-2000Revision 1.0Date MAY 1996Page 7 of 7RECHARGE If the battery voltage is less than 11.76 volts, connect theBATTERY datalogger to the bat
QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF THE HANDAR 540A/570A DCPTYPETECHNICAL INSTRUCTIONNUMBER4250-20
Number 4250-2010Revision 0.1Date MAY 1996Page i of iiTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONSIBILITIES 12.1 Project Man
Number 4250-2010Revision 0.1Date MAY 1996Page ii of iiLIST OF FIGURES (CONTINUED)Figure Page4-2 Front Panel Configurations - Handar 540A/570A DCP 54-
Number 4250 Revision 1.0 Date MAY 1996 Page i of iTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONS
Number 4250-2010Revision 0.1Date MAY 1996Page 1 of 371.0 PURPOSE AND APPLICABILITYThis technical instruction (TI) describes procedures for servicing a
Number 4250-2010Revision 0.1Date MAY 1996Page 2 of 37• Coordinate with Handar for return, servicing, and recalibration of 540A/570A DCPsthat fail to o
Number 4250-2010Revision 0.1Date MAY 1996Page 3 of 37• Electronic contacts cleaning fluid• Medium screwdrivers (flat-blade and Phillips-head)4.0 METHO
Number 4250-2010Revision 0.1Date MAY 1996Page 4 of 37POST-FIELD INSPECTION CHECKLISTHANDAR 540A/570A DCPDCP S/N: ______________________________Site:
Number 4250-2010Revision 0.1Date MAY 1996Page 5 of 37Figure 4-2. Front Panel Configurations - Handar 540A/570A DCP.
Number 4250-2010Revision 0.1Date MAY 1996Page 6 of 374.1.2 Physical Inspection - ExternalINSPECT Inspect the outside of the DCP case thoroughly. Look
Number 4250-2010Revision 0.1Date MAY 1996Page 7 of 37Figure 4-3. DCP Component Diagram.
Number 4250-2010Revision 0.1Date MAY 1996Page 8 of 37SWITCH When the "TERM" setup screen (see Figure 4-4) is displayed onBAUD RATE the compu
Number 4250-2010Revision 0.1Date MAY 1996Page 9 of 37Figure 4-4. TERM Setup Screen.
Number 4250-2010Revision 0.1Date MAY 1996Page 10 of 37Figure 4-5. "TERM" DCP Programming Screen (Program in DCP Memory).
Number 4250 Revision 1.0 Date MAY 1996 Page 1 of 51.0 PURPOSE AND APPLICABILITYThis standard operating procedure (SOP) ou
Number 4250-2010Revision 0.1Date MAY 1996Page 11 of 37Figure 4-6. TERM DCP Programming Screen (No Program in DCP Memory).
Number 4250-2010Revision 0.1Date MAY 1996Page 12 of 37SWITCH Switch the DCP from "RUN" mode to "PROGRAM" mode byMODES entering ?.T
Number 4250-2010Revision 0.1Date MAY 1996Page 13 of 37COMPARE The DCP data channel 04 reading (Rotronics AT output signal)READINGS must be adjusted by
Number 4250-2010Revision 0.1Date MAY 1996Page 14 of 37ROUTINE SERVICING CHECKLISTHANDAR 540A/570A DCPDCP S/N: __________________________Date: ________
Number 4250-2010Revision 0.1Date MAY 1996Page 15 of 37VERIFY Verify that the GOES primary channel select switches are set toGOES “900” and the seconda
Number 4250-2010Revision 0.1Date MAY 1996Page 16 of 37CLEAN INTERIOR Clean the inside of the DCP with compressed air.OF DCPMODIFY 570A If this is a Ha
Number 4250-2010Revision 0.1Date MAY 1996Page 17 of 37Figure 4-8. Handar 570A ADC Board - Component Locations.
Number 4250-2010Revision 0.1Date MAY 1996Page 18 of 37Figure 4-9. Handar 540A Met Board - Component Locations.
Number 4250-2010Revision 0.1Date MAY 1996Page 19 of 374.3 DCP PROGRAMMINGThe Handar 540A/570A DCP operational configuration is established through a u
Number 4250-2010Revision 0.1Date MAY 1996Page 20 of 37Table 4-1DCP ID AssignmentsIMPROVE Transmissometer NetworkDCP-ID CHAN LOCATION TIME * RATE **FA4
Number 4250 Revision 1.0 Date MAY 1996 Page 2 of 52.2 INSTRUMENT TECHNICIANThe instrument technician shall:• Perform all
Number 4250-2010Revision 0.1Date MAY 1996Page 21 of 37EXECUTE Execute the Handar DCP communications program "TERM" fromPROGRAM the PC comput
Number 4250-2010Revision 0.1Date MAY 1996Page 22 of 37• $ Forced Scan• # Forced Transmit (GOES Radio)Enter I to edit the station ID. The download pro
Number 4250-2010Revision 0.1Date MAY 1996Page 23 of 37Enter M to select DCP channel 01. The display message will be "PCHANNEL 01."Enter N to
Number 4250-2010Revision 0.1Date MAY 1996Page 24 of 37LABORATORY PERFORMANCE TESTINGHANDAR 540A/570A DCPDCP S/N: __________________________Date: _____
Number 4250-2010Revision 0.1Date MAY 1996Page 25 of 374.4.3 Run Mode Timing ChecksVERIFY Verify that the GOES primary channel select switches are set
Number 4250-2010Revision 0.1Date MAY 1996Page 26 of 37Set the power element of the wattmeter for the forward direction.Monitor the wattmeter reading a
Number 4250-2010Revision 0.1Date MAY 1996Page 27 of 37APPENDIX AHANDAR 540A DCP CONFIGURATION PROGRAM - 540ROT.DCP
Number 4250-2010Revision 0.1Date MAY 1996Page 28 of 37P ID FA43D0BEP STATION TIME 15:47:38P YEAR (XX) 94P DCP JULIAN DATE 055P GOES PRI XMT
Number 4250-2010Revision 0.1Date MAY 1996Page 29 of 37P CHANNEL NO. 03P03 SENSOR TYPE 10P03 CARD SLOT # 06
Number 4250-2010Revision 0.1Date MAY 1996Page 30 of 37P CHANNEL NO. 06P06 SENSOR TYPE 10P06 CARD SLOT # 06
Number 4250 Revision 1.0 Date MAY 1996 Page 3 of 5• Reference thermometer (°C)• Replacement components as required• Batte
Number 4250-2010Revision 0.1Date MAY 1996Page 31 of 37P08 LOW DIFF LIMIT NO LIMITP CHANNEL NO. 09P09 SENSOR TYPE 10
Number 4250-2010Revision 0.1Date MAY 1996Page 32 of 37APPENDIX BHANDAR 570A DCP CONFIGURATION PROGRAM - 570TROT.DCP
Number 4250-2010Revision 0.1Date MAY 1996Page 33 of 37P ID FA43D0BEP STATION TIME 22:46:25P YEAR (XX) 94
Number 4250-2010Revision 0.1Date MAY 1996Page 34 of 37P02 MEAS INTERVAL 01:00:00P02 START OF MEAS 17:30:00P02 LEVEL 1 MEAS TYPE 001
Number 4250-2010Revision 0.1Date MAY 1996Page 35 of 37P05 SENSOR TYPE 10P05 SENSOR NAME TAG 10P05 CARD SLOT # 01
Number 4250-2010Revision 0.1Date MAY 1996Page 36 of 37P07 ZERO SCALE 000P07 MEAS INTERVAL 01:00:00P07 START OF MEAS 17:30:00
Number 4250-2010Revision 0.1Date MAY 1996Page 37 of 37P CHANNEL NO. 10P10 SENSOR TYPE 12P10 SENSOR NAME TAG 12
Number 4250 Revision 1.0 Date MAY 1996 Page 4 of 5• Frequency counter• Standard electronics laboratory small tools• Solte
Number 4250 Revision 1.0 Date MAY 1996 Page 5 of 5• Routine laboratory servicing• DCP programming• Pre-field performance
QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF CAMPBELL 21X DATALOGGERSTYPETECHNICAL INSTRUCTIONNUMBER4250-20
Number 4250-2000Revision 1.0Date MAY 1996Page i of iTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONSIBILITIES 12.1 Project Mana