Appendix B. Correcting Electrical Conductivity Measurements for System Losses
if Shorted=high then
TDR
For I=1 To 20
vector(I)=wave(I+9)
Next
AvgSpa (rho(2),20,vector(1))
Shorted=low
endif
'write results to output storage
If WriteToOutput=high Then
CallTable rhoTable
WriteToOutput=low
EndIf
'setting SDMports high will configure control ports 1 thru 3 to allow connection
'of TDR100 to PC using PCTDR
If SDMports=high Then
PortsConfig (&B00000111,&B00000000)
SDMports=low
EndIf
NextScan
EndProg