
Appendix A. Discussion of TDR Probe Offset and a Simple Laboratory Method for Calculation
A.3 Method for Calculating Probe Offset Using
Information from the Terminal Mode of PCTDR
Letting V
p
= 1 and solving [1] for ProbeOffset gives
LastartendProbeOff −
[A3]
The start and end distance values are determined by an algorithm in the
TDR100 operating system. The apparent length, La, is related to actual rod
length and permittivity by
)
TL:La ε⋅=
[A4]
The rod length, L, is the physical length of the rods (m). For 3-rod TDR
probes having longer outer rods the length of the outer rods is used.
The dielectric permittivity of water can be calculated from water temperature
using
[A5]
() () () (
[]
3
8
2
53
25T108.225T1019.125T105791.4154.78:Tε −⋅⋅−−⋅⋅+−⋅⋅−⋅=
−−−
)
Table 1 contains dielectric permittivities for a typical range of temperatures and
may be used in lieu of equation [A5]. Substituting the calculation of La using
equations [A4] and [A5] into equation [A3] leaves the end and start distances
as the only unknowns. These values can be acquired directly from the TDR100
algorithm by using the terminal emulator mode of PCTDR.
A.3.1 Procedure for Calculating Probe Offset
Connect all cabling and multiplexers to be used for field or laboratory
measurements.
Immerse the TDR probe rods in DI or tap water. The container must be large
enough so rods are at least 5 cm from container walls.
Use PCTDR as follows.
1. Using settings/calibration functions/volumetric water content , select La/L
as the output by choosing the upper User Defined equation and setting
slope to 1 and offset to 0.
2. Enter values for Waveform parameters. Suggested values are: Average =
4, Points = 251. Relative propagation velocity, Vp, must be 1. Choice of
waveform length depends on the actual probe rod length. There should be
about .5 meters before the probe, enough distance for probe apparent
length in water (approximately 9 times rod length, and enough distance for
the waveform past the end of the probe. The distance for the end can be
approximated as 3 times rod length.
3. Enter the value for probe rod length in meters and set probe offset to 0 m.
4. Click the Water Content button to send the values to the TDR100 and to
have it calculate La/L and provide start and end values.
A-3
Commenti su questo manuale