CSAT3 Three Dimensional Sonic Anemometer
32
17: Real Time (P77)
1: 110 Day,Hour/Minute (midnight = 0000)
18: Resolution (P78)
1: 1 High Resolution
19: Sample (P70)
1: 3 Reps
2: 13 Loc [ wnd_spd_a ]
20: Sample (P70)
1: 1 Reps
2: 17 Loc [ csat_dir ]
21: Average (P71)
1: 4 Reps
2: 1 Loc [ Ux ]
22: Standard Deviation (P82)
1: 4 Reps
2: 1 Sample Loc [ Ux ]
23: Totalize (P72)
1: 7 Reps
2: 20 Loc [ smpls ]
*Table 2 Program
02: 0 Execution Interval (seconds)
*Table 3 Subroutines
;Configure CSAT3.
;
1: Beginning of Subroutine (P85)
1: 1 Subroutine 1
;Check the diagnostic word for lost trigger condition.
;Set Flag 1 low to reconfigure the CSAT3 if it is in a
;lost trigger state.
;
2: If (X<=>F) (P89)
1: 5 X Loc [ diag ]
2: 1 =
3: 61440 F
4: 21 Set Flag 1 Low
;Configure the CSAT3. This is done the first time
;through Table 1 or if the CSAT3 loses the trigger.
;
3: If Flag/Port (P91)
1: 21 Do if Flag 1 is Low
2: 30 Then Do