CNR4 Net Radiometer
FIGURE 6-2. The marks on the end of the CNR4: S for SOLAR cable,
and T for TEMP cable.
The measurement details for Pt-100 sensor, including the wiring diagram and
sample program are explained in the Appendix C of this manual.
The four radiation outputs can be measured using differential or single-ended
inputs on the datalogger. A differential voltage measurement is recommended
because it has better noise rejection than a single-ended measurement.
When differential inputs are used, jumper the low side of the
input to AG or to keep the signal in common mode range.
NOTE
The Tables 6-1 and 6-2 show the wiring instructions for the differential
measurement and single-ended measurement connections to the datalogger,
respectively. The cables have the white band at the pigtail end of the cable
with the color keys. See the Figure 6-3 and 6-4 below for the labels on the
cable for both the SOLAR and TEMP cables.
FIGURE 6-3. Labels on the pigtail end of the SOLAR cable.