Pagina 1 - CNR4 Net Radiometer

CNR4 Net RadiometerRevision: 11/10Copyright © 2000-2010Campbell Scientific, Inc.

Pagina 2 - Warranty and Assistance

CNR4 Net Radiometer 2.2 Pyranometer Specifications * indicates ISO specifications. Spectral range: 305 to 2800 nm (50% points) Sensitivity: 10 to 20

Pagina 3 - CNR4 Table of Contents

CNR4 Net Radiometer 2.3 Pyrgeometer Specifications Spectral range: 4.5 μm to 42 μm (50% points) Sensitivity: 5 to 15 μV/W/m2 Impedance: 20 Ω to 200

Pagina 4

CNR4 Net Radiometer 2.4.1 CNF4 Specifications Heater Power consumption: 10 W @ 12 Vdc (15 Ω) Ventilator Power consumption: Supply voltage: 5 W @ 12

Pagina 5

CNR4 Net Radiometer Do not attempt to rotate the instrument using the sensor heads, or you may damage the sensors; use the mounting rod only. NOTE 4.

Pagina 6

CNR4 Net Radiometer FIGURE 3-2. Attaching the CNR4 onto the mounting rod (CSI p/n 26120) using vertical pole or horizontal crossarm. For installat

Pagina 7

CNR4 Net Radiometer 5. Using the CNR4 in the Four Separate Components Mode In the four separate components mode configuration (measuring two short-wa

Pagina 8 - 2. Sensor Specifications

CNR4 Net Radiometer colder. This means that for estimating the far infrared radiation that is generated by the object that is faced by the pyrgeomete

Pagina 9 - 2.1 CNR4 Specifications

CNR4 Net Radiometer The thermistor measurement can be done by the datalogger, using the half bridge measurement method which requires one voltage exci

Pagina 10

CNR4 Net Radiometer TABLE 5-2. Resistance values versus CNR4's Pt-100 temperature in °C. Temperature [°C] Resistance[Ω] Temperature[°C] Resista

Pagina 11 - < 6 W/m

CNR4 Net Radiometer In the equation above, E is calculated according to the Equation 5-1. Albedo will always be smaller than 1. Checking this can be

Pagina 12 - 3. Installation

Warranty and Assistance The CNR4 NET RADIOMETER is warranted by Campbell Scientific, Inc. to be free from defects in materials and workmanship under n

Pagina 13

CNR4 Net Radiometer 5.7 Calculation of Net (Total) Radiation In the four separate components mode, net radiation, Rn, can be calculated using the ind

Pagina 14

CNR4 Net Radiometer FIGURE 6-2. The marks on the end of the CNR4: S for SOLAR cable, and T for TEMP cable. The measurement details for Pt-100 sensor

Pagina 15 - Pyrgeometer

CNR4 Net Radiometer FIGURE 6-4. Labels on the pigtail end of the TEMP cable. TABLE 6-1. Datalogger Connections for Differential Measurement Functio

Pagina 16

CNR4 Net Radiometer 7. Datalogger Programming The CNR4 outputs four voltages that typically range from 0 to 15 mV for the pyranometers, and ± 5 mV fo

Pagina 17

CNR4 Net Radiometer Average Corrected long-wave radiation (pyrgeometer up) Average Corrected long-wave radiation (pyrgeometer down) Average Shor

Pagina 18 - 5.4 Calculation of Albedo

CNR4 Net Radiometer Public cnr4_T_C 'CNR4 thermistor temperature in Celcius Public cnr4_T_K 'CNR4 thermistor temp

Pagina 19

CNR4 Net Radiometer Average (1,Rl_net,IEEE4,False) Average (1,albedo,IEEE4,False) Average (1,Rn,IEEE4,False) EndTable DataTable (cnr4_ts,True,-1)

Pagina 20 - 6. Wiring

CNR4 Net Radiometer 7.2.2 Example 2, CR3000 Program Using Differential Measurements The Program Example 2 requires four differential channels to meas

Pagina 21

CNR4 Net Radiometer '8H '8L CNR4 thermistor signal (white) 'gnd CNR4 thermistor signal reference (black) '

Pagina 22

CNR4 Net Radiometer 'CNR4 multipliers Public cnr4_mult(4) Const pyranometer_up_mult = 1000/pyranometer_up_sensitivity '(W/m^2/mV) Const pyr

Pagina 23 - 7. Datalogger Programming

CNR4 Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for

Pagina 24

CNR4 Net Radiometer 'Compute short-wave net radiation Rs_net = short_up - short_dn 'Compute long-wave net radiati

Pagina 25

CNR4 Net Radiometer 'CR5000 Series Datalogger ' 'CNR4 program 'This program measures CNR4 four-component net radiometer 'This

Pagina 26

CNR4 Net Radiometer Public Rl_net 'long-wave net radiation Public albedo 'Albedo Public Rn 'total net radiation Units

Pagina 27

CNR4 Net Radiometer DataTable (cnr4_ts,True,-1) DataInterval (0,1,Sec,10) CardOut (1,-1) Sample (4,cnr4(1),IEEE4) Sample (1,cnr4_T_K,IEEE4) End

Pagina 28

CNR4 Net Radiometer 8. Troubleshooting If there is no indication as to what may be the problem, start performing the following "upside-down test

Pagina 29

CNR4 Net Radiometer 8.2 Testing the Pyrgeometer It is assumed that the zero offset is no more than a few watts per square meter (see second test in s

Pagina 30

CNR4 Net Radiometer 9.2 Recalibration For quality assurance of the measured data, the manufacturer recommends the CNR4 be recalibrated on a regular s

Pagina 31

CNR4 Net Radiometer 9.4 Replacement Parts The following is the list of replacement parts for the CNR4 and CNF4 (heater/ventilator) available from Cam

Pagina 32

CNR4 Net Radiometer 32

Pagina 33

Appendix A. CNR4 Performance and Measurements under Different Conditions Below, Table A-1 shows what one might typically expect to measure under diff

Pagina 34 - 8. Troubleshooting

CNR4 Table of Contents 9. Maintenance and Recalibration ...29 9.1 Cleaning Windows and Domes...

Pagina 35 - 8.4 Testing the Pt-100

Appendix A. CNR4 Performance and Measurements under Different Conditions FIGURE A-1. Different measurement conditions and signals. Upperpyrgeomete

Pagina 36 - 9.2 Recalibration

Appendix A. CNR4 Performance and Measurements under Different Conditions upwellingsignal(downwardfacing)pyrgeometer‐30‐20‐1001020300:00:001:00:00

Pagina 37 - 9.4 Replacement Parts

Appendix A. CNR4 Performance and Measurements under Different Conditions 4/181067.5⎥⎦⎤⎢⎣⎡⋅=−CG3upper EatureSky temper (B-1) 4/181067.5⎥⎦⎤⎢⎣⎡⋅=−CG3

Pagina 38

Appendix B. CNF4 Heater/Ventilator Whenever the heater is used, the heating may cause errors in the measurement of the sensor temperature. Under most

Pagina 39 - Conditions

Appendix B. CNF4 Heater/Ventilator 10 Watt power available? Not available DO NOT HEAT Available Consider options below Clock and relay available? N

Pagina 40 - Upperpyrgeometer

Appendix B. CNF4 Heater/Ventilator B.2 Attaching the Optional CNF4 Heater/Ventilator Unit to CNR4 1. The CNF4 heater/ventilator unit comes with the

Pagina 41

Appendix B. CNF4 Heater/Ventilator 2. Attach the heater/ventilator unit unto the bottom of the CNR4 sensor, using the three pan-head screws and wash

Pagina 42

Appendix B. CNF4 Heater/Ventilator 3. Make sure the cables are cleared from the edges of the CNF4, as shown in Figure B-3, and place the white solar

Pagina 43 - B.1 General Information

Appendix B. CNF4 Heater/Ventilator FIGURE B-5. Attaching the solar shield to CNF4 using four flat-head screws. 4. Once the CNF4 heater/ventilator

Pagina 44

Appendix B. CNF4 Heater/Ventilator B.3 Wiring The following table shows the recommended datalogger wiring for using the CNR4 sensor with the CNF4 he

Pagina 45 - Unit to CNR4

CNR4 Table of Contents Tables 5-1. Resistance values versus CNR4's thermistor temperature in °C ...11 5-2. Resistance values versus CNR

Pagina 46

Appendix B. CNF4 Heater/Ventilator B.4 Example B, CR3000 Datalogger Program with Heater/Ventilator Control The Program Example B measures the four r

Pagina 47

Appendix B. CNF4 Heater/Ventilator ' ' '8H '8L CNR4 thermistor signal (white) 'gnd CNR4 thermistor signal refe

Pagina 48

Appendix B. CNF4 Heater/Ventilator Dim Rs, Vs_Vx 'CNR4 sensitivities: refer to the Certificate of Calibration from Kipp & Zonen for sensiti

Pagina 49 - B.3 Wiring

Appendix B. CNF4 Heater/Ventilator 'CNR4 thermistor measurement BrHalf (Vs_Vx,1,mv5000,16,Vx1,1,2500,True ,0,250,1.0,0) Rs =

Pagina 50 - Heater/Ventilator Control

Appendix B. CNF4 Heater/Ventilator B-12 pulling it straight out. Inspect the fan inside for any object that might impede the fan’s rotation. Upon c

Pagina 51

Appendix C. CR3000 Program for Measuring Pt-100 Temperature Sensor The program example C measures the Pt-100 sensor for the body temperature of the C

Pagina 52

Appendix C. CR3000 Program for Measuring Pt-100 Temperature Sensor 'CR3000 Series Datalogger ' 'CNR4 program 'This program measur

Pagina 53 - B.5.1 Testing the Heater

Appendix C. CR3000 Program for Measuring Pt-100 Temperature Sensor Public cnr4_T_C 'CNR4 thermistor temperature in Celcius Public cnr4_T_K &

Pagina 54

Appendix C. CR3000 Program for Measuring Pt-100 Temperature Sensor C-4 Average (1,albedo,IEEE4,False) Average (1,Rn,IEEE4,False) EndTable DataTabl

Pagina 56

CNR4 Table of Contents iv

Pagina 57

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Pagina 58

CNR4 Net Radiometer 1. General Description The CNR4 is a four component net radiometer that measures the energy balance between incoming and outgoing

Pagina 59

CNR4 Net Radiometer 2. Sensor Specifications The CNR4 consists of two pyranometers, for measuring short-wave radiation, and of two pyrgeometers for m

Pagina 60

CNR4 Net Radiometer 2.1 CNR4 Specifications Sensor sensitivities: Four probes have unique sensitivity values. Please refer to the calibration sheets