CS526 ISFET pH Probe
7.2.3.2 SerialOut() Instruction
The SerialOut() instruction has the following syntax:
SerialOut(ComPort,OutString,WaitString,NumberTries,TimeOut)
ComPort — the datalogger COM port in which the probe is connected.
OutString — use "1M1!"+CHR(13) for the OutString when the default probe
address of 1 is used.
WaitString — enter the null (“”) WaitString to tell the datalogger to wait for
the echo of each character in the OutString
NumberTries — enter 0
TimeOut — specifies the time, in 0.01 seconds, that the datalogger should
wait for the WaitString or echo of each character in the OutString (0 is
used in the example program).
7.2.3.3 SerialIn() Instruction
The SerialIn() instruction has the following syntax:
SerialIn(Dest,ComPort,TimeOut,TerminationChar,MaxNumChars)
Dest — specifies the variable in which the incoming data will be stored.
Please note that in the beginning of the CRBasic program, this variable
needs to be declared as ASCII string format (see example program in
Appendix B, Example Program)
ComPort — the datalogger COM port in which the probe is connected.
TimeOut — 20 should be adequate, which gives a 200 ms maximum delay
time. The TimeOut parameter is used to specify the amount of time, in
0.01 seconds, that the datalogger should wait before proceeding to the
next instruction.
TerminationChar — enter 0
MaxNumChars — 20 should be adequate (specify the maximum number of
characters to expect per input)
8