CS526 ISFET pH Probe
1. Introduction
The CS526 ISFET pH Probe measures pH from 0 to 14 in aqueous or semi-
solid solutions. It outputs TTL serial data that is read by compatible
dataloggers (see Section 6 Specifications).
Before using the CS526, please study
Section 2. Cautionary Statements
Section 3. Initial Inspection
Section 4. Quickstart
More detailed instructions for operation, troubleshooting, and maintenance are
available in the remaining sections.
2. Cautionary Statements
Maximum input voltage is 5 Vdc. Incorrect wiring may cause loss of
performance and irreversible damage.
To prevent scratching the sensor chip when cleaning, first soak the sensor in
soapy water, then gently scrub with a toothbrush (see Section 11.
Maintenance). Most scratches occur when hard particles are rubbed on the
chip surface. Scratches cause irreversible damage to the probe.
The CS526 is rugged, but it should be handled as a precision scientific
instrument.
The CS526 has no user-serviceable parts. Any attempt to disassemble the
device will void the six-month warranty.
Care should be taken when opening the shipping package to not damage or cut
the cable jacket. If damage to the cable is suspected, consult with a Campbell
Scientific applications engineer.
3. Initial Inspection
Upon receipt of the CS526, inspect the packaging and contents for damage.
File damage claims with the shipping company.
The model number and cable length are printed on a label at the connection
end of the cable. Check this information against the shipping documents to
ensure the correct product and cable length are received.