CS526 ISFET pH Probe
Maximum Cable Length: 100 m (328 ft)
Cable Type: Three-twisted pair, 24 awg cable with
Santoprene
®
jacket
Sensor Material: PEEK
Weight
w/10 ft Cable: 318 g (11.2 oz)
Dimensions
Length: 102 mm (4 in)
Diameter: 16 mm (0.63 in)
Certifications: ISO 9001
compliant
EMC standard IEC61326:2005
7. Installation and Wiring
7.1 Preparation for Use
The CS526 is shipped dry. Before the probe can be calibrated, soak it in 7 pH
buffer solution for 15 minutes. Unlike many glass-bulb type pH sensors, the
CS526 ISFET pH probe can be installed without regard to orientation.
7.2 Wiring
TABLE 7-1. CR800, CR850, CR1000, and CR3000 Wiring
Wire Color Wire Label/Function Datalogger Connections
Red (see following
caution)
Power 5V 5V
Black Ground G
White Signal #1 (Tx) Control Port (Tx)
Green Signal #2 (Rx) Control Port (Rx)
Clear Shield G
This probe must be connected to the datalogger’s 5V
terminal (not 12V). Connecting to a higher voltage will
damage the probe beyond repair.
CAUTION
8. Programming
8.1 Get Data Command
The datalogger needs to send a “get data” serial command to the CS526 to get
the pH data. This command is sent to the CS526 via the SerialOut() CRBasic
4