
CS106 Barometric Pressure Sensor
9. Maintenance and Calibration
Since the sensor is semi-sealed, minimum maintenance is required:
1. Visually inspect the cable connection to ensure it is clean and dry.
2. Visually inspect the casing for damage.
3. Ensure that the pneumatic connection and pipe are secure and undamaged.
The external case can be cleaned with a damp, lint-free cloth and a mild
detergent solution.
Vaisala recommends recalibration every two years under normal use. In areas
where a lot of contaminants are present, recalibration every year is
recommended.
Contact Campbell Scientific, Inc. (435-227-9000) for an RMA number before
returning the sensor for recalibration.
Should you lose the five terminal connector (pn 16004), the replacement part
can be purchased from Campbell Scientific, Inc. Contact Campbell Scientific,
Inc. to purchase the part.
The CS106 is sensitive to static when the backplate is
removed. To avoid damage, take adequate anti-static
measures when handling.
13
Commenti su questo manuale