CR10 and VW Sensors, 2002/12/18 5
Sensors and Sweeps
Sensor Table
Find your sensor and its required sweep. You can also find
sweeps and usage in the sensor manual. After you find the
sweep, look at the Sweep Table below.
Sweep Table
Each row lists a sweep and the values needed for instruction 28
and 131. (Instruction 28 does not use T or N). If your sensor is
not listed above, check your sensor calibration sheet to find the
lowest and highest frequencies in the calibration, and then
choose the sweep that is closest to those frequencies.
Name Part Number Required Sweep
Crackmeter 5263608x Sweep 3 or 2
Displacement Sensor , Extensometer 526363xx Sweep 3 or 2
Piezometer 526110xx, 526210xx Sweep 3
Settlement Cell, 50 or 100 psi 52612020, 52612030 Sweep 3
Strain Gauge, Arc-Weldable 52630306 Sweep 2 or 1
Strain Gauge, Embedment 52640126 Sweep 2 or 1
Strain Gauge, Spot-Weldable 5260210x Sweep 2 or 3
Stress Station, VW Transducers 526081xx, 526114xx Sweep 3
Total Pressure Cell 526082xx,5260828x Sweep 3
Total Pressure Cell, Radial 5260826x Sweep 3
Total Pressure Cell, Tangential 5260827x Sweep 3
Starting Freq Ending Freq T N Cycles
Sweep 1* 450 1125 70 40 100
Sweep 2* 800 2000 22 26 100
Sweep 3 1400 3500 14 32 250
Sweep 4 2300 6000 8 27 500
* These sweeps are not used with instruction 28.