CS451/CS456 Submersible Pressure Transducer
7.1.1.3 Powering the Sensor
The A200 provides power to the sensor when it is connected to a PC’s USB
port. An internal DC/DC converter boosts the 5 Vdc supply from the USB
connection to a 12 Vdc output is required to power the sensor.
7.1.1.4 Determining which COM Port the A200 has been Assigned
When the A200 driver is loaded, the A200 is assigned a COM port number.
This COM port number is needed when using the Device Configuration Utility
or a PC terminal software such as HyperTerminal.
Often, the assigned COM port will be the next port number that is free.
However, if other devices have been installed in the past (some of which may
no longer be plugged in), the A200 may be assigned a higher COM port
number.
To check which COM port has been assigned to the A200, watch for the
appearance of a new COM port in the list of COM ports offered in the software
package (e.g., LoggerNet) before and after the installation, or look in the
Windows Device Manager list under the ports section (access via the control
panel).
FIGURE 7-1. A200 Sensor-to-PC Interface
TABLE 7-2. A200 Wiring
Color
Sensor
Cable Label
A200
Terminal
Red 12V +12Vdc
Black G G
White C Tx
Blue G Rx
Yellow G G
Clear Signal Ground G
11