Campbell TDR100 Manuale delle Istruzioni Pagina 40

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 56
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 39
TDR100
34
;Measure Electrical Conductivity on the same CS610 TDR probe connected to Channel 3.
9: TDR100 Measurement (P119)
1: 0 SDM Address
2: 3 Electrical Conductivity
3: 3001 MMMP Mux & Probe Selection
4: 4 Waveform Averaging
5: 1 Vp
6: 250 Points
7: 9.5 Cable Length (meters)
8: 5 Window Length (meters)
9: .3 Probe Length (meters)
10: .085 Probe Offset (meters)
11: 5 Loc [ EC_CH3 ]
12: 7.4 Mult
13: 0.0 Offset
;Get the raw waveform for the CS610 probe on Channel 3.
;When creating this program in Edlog, the user must reserve 259 input locations
;for data storage. How? In Edlog click on
;"Edit", select "Input Labels". In the Input Location Editor click on "Edit", select
;"Insert Block", and identify 259 input storage location for the captured waveform.
;Note: 9 parameter elements + 250 waveform data points = 259 input location/waveform.
10: TDR100 Measurement (P119)
1: 0 SDM Address
2: 1 Waveform (256 locs)
3: 3001 MMMP Mux & Probe Selection
4: 4 Waveform Averaging
5: 1 Vp
6: 250 Points
7: 9.5 Cable Length (meters)
8: 5 Window Length (meters)
9: .3 Probe Length (meters)
10: .085 Probe Offset (meters)
11: 11 Loc [ WF_1 ]
12: 1.0 Mult
13: 0.0 Offset
;Turn off the switched 12V to power off the TDR100:
11: Do (P86)
1: 55 Set Port 5 Low
;Set the Output Flag to output data each time measurements are made:
12: Do (P86)
1: 10 Set Output Flag High (Flag 0)
;Output a time stamp; year, day, and hour/minute:
13: Real Time (P77)
1: 1220 Year,Day,Hour/Minute (midnight = 2400)
Vedere la pagina 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 55 56

Commenti su questo manuale