IRGASON Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
Scrubber bottle cover
Scrubber bottle cover
FIGURE 9-5. Replacing the desiccant and CO2 scrubber bottles
9.6 Factory Recalibration
When the IRGASON is manufactured, the gas analyzer calibrated over a wide
range of temperatures, pressures, and gas concentrations. All CO
2
calibration
gases used in this process are mixtures of CO
2
in ambient air and are traceable
to the WMO Mole Fraction Scale maintained by the Central Carbon Dioxide
Laboratory and the Carbon Cycle Greenhouse Gases Group of the Global
Monitoring Division / National Oceanographic and Atmospheric
Administration in Boulder, CO, USA.
After an extended period of time in the field, the IRGASON gas analyzer may
need to undergo this factory calibration again to ensure valid measurements.
When recalibration is deemed necessary, contact Campbell Scientific.
The IRGASON sonic anemometer also undergoes an initial factory calibration.
Usually this calibration remains valid unless a transducer fails or damage to the
instrument leads to a change in geometry. The sonic anemometer requires
calibration when
it develops a wind offset greater than the specification.
it sets diagnostic flags under dry conditions with little to no wind and
with no obstruction in the ultra sonic paths.
The wind offset is tested by creating a zero wind environment by encircling the
sensor with a large plastic bag (unused refuse bag), using caution not to block
the sonic paths. See the manual CSAT3 Three Dimensional Sonic Anemometer,
Section 11.2.2, Test for Wind Offset, available at www.campbellsci.com
.
When recalibration is deemed necessary, contact Campbell Scientific.