Campbell HMP 35A Manuale delle Istruzioni Pagina 20

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 19
HMP155ATemperature and Relative Humidity Probe
Connections to CRBasic dataloggers are given in TABLE 7-1 and TABLE 7-2.
To wire an Edlog datalogger, see an older manual at
www.campbellsci.com/old-manuals, or contact a Campbell Scientific
application engineer for assistance.
When measuring the HMP155A with single-ended
measurements, the white and black leads must both be
connected to
. Doing otherwise will connect the
datalogger’s analog and power ground planes to each other,
which in some cases can cause offsets on low-level analog
measurements. To avoid 3 mA flowing into analog ground,
switch the sensor on/off for its own measurement.
TABLE 7-1. Connections for Single-Ended Measurements
Color
Wire Label
CR6
CR9000X
CR1000,
CR3000,
CR800,
CR5000
CR200(X)
Yellow Temp Signal Single-Ended Input Single-Ended Input
Blue RH Signal Single-Ended Input Single-Ended Input
White Signal Reference
Black Signal Ground
Clear Shield
Red Power SW12V SW12V SW Power
TABLE 7-2. Connections for Differential Measurements
Color
Wire Label or
Description
CR6
CR1000,
CR3000,
CR800,
CR5000,
CR9000X
Yellow Temp Signal Differential Input – H
Jumper to White Differential Input – L
Blue RH Signal Differential Input – H
White Signal Reference Differential Input – L
Black Signal Ground G
Clear Shield
Red Power SW12V 12V/SW12V
CAUTION
12
Vedere la pagina 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 43 44

Commenti su questo manuale

Nessun commento