Campbell CURS100 Manuale Utente Pagina 11

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 18
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 10
User Manual
3
3.1 Differential Measurement
The voltage across the completion resistor is measured with the differential
voltage measurement. Use VoltDiff() for the CRBasic dataloggers (for example,
CR1000, CR5000, or CR9000(X)). Use Instruction 2 for Edlog dataloggers (for
example, CR10X). The differential voltage measurement measures the difference
in voltage between the low and high terminals. The CURS100 connects the
resistor between the high and the low terminals.
3.2 Completing the Current Loop Circuit
As shown in Figure 2-1, the 100 Ω sense resistor in the CURS100 is not connected
to the adjacent ground pin that connects into the datalogger signal ground ( or
AG). Hence, an additional connection must be made in order to complete the
loop. Which is commonly done by connecting the CURS100 L terminal to a
datalogger G (power ground) terminal with a jumper wire (Figure 3-1).
Connecting the L terminal to the adjacent ground ( or G) terminal on the
CURS100 will result in unwanted return currents flowing into the datalogger
signal ground ( or AG) which could induce undesirable offset errors in low-
level, single-ended measurements. The ground ( or G) terminal on the
CURS100 can be used to connect cable shields to ground.
Completing the loop by connecting voltages other than ground is possible as long
as the datalogger voltage input limits are not exceeded. These input limits specify
the voltage range, relative to datalogger ground, which both H and L input
voltages must be within in order to be processed correctly by the datalogger. The
input limits are ±2.5 V for the CR10(X) and ±5 V for the CR800, CR850,
CR1000, CR3000, CR5000, and CR9000(X). Hence, when measuring currents up
to 50 mA with the CURS100, a connection to datalogger ground is necessary in
order for the resulting (50 mA) (100 Ω) = 5 V signal to comply with the ±5 V
input limits for the CR800, CR850, CR1000, CR3000, CR5000, and CR9000(X)
dataloggers.
Vedere la pagina 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 17 18

Commenti su questo manuale

Nessun commento