Campbell CS650 Manuale delle Istruzioni

Navigare online o scaricare Manuale delle Istruzioni per Misurazione, test & controllo Campbell CS650. Campbell CS650 Instruction manual [en] Manuale Utente

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa

Sommario

Pagina 1 - INSTRUCTION MANUAL

INSTRUCTION MANUAL CS650 and CS655 Water Content Reflectometers Revision: 5/15 Copyright © 2011- 2015 Campbell Scientific,

Pagina 2

Table of Contents iv

Pagina 3 - Limited Warranty

CS650 and CS655 Water Content Reflectometers 1. Introduction The CS650 and CS655 are multiparameter smart sensors that use innovative techniques to

Pagina 4 - Assistance

CS650 and CS655 Water Content Reflectometers 4. Quickstart Short Cut is an easy way to program your datalogger to measure the CS650 and assign datal

Pagina 5 - Precautions

CS650 and CS655 Water Content Reflectometers 4. Select Datalogger Model and Scan Interval (default of 5 seconds is OK for most applications). Click

Pagina 6

CS650 and CS655 Water Content Reflectometers 6. After selecting the sensor, click at the left of the screen on Wiring Diagram to see how the sensor

Pagina 7 - Table of Contents

CS650 and CS655 Water Content Reflectometers FIGURE 5-1. CS650 Water Content Reflectometer Volumetric water content information is derived from the

Pagina 8

CS650 and CS655 Water Content Reflectometers means of RS-232 Tx/Rx. The A200 USB-to-Serial Module allows RS-232 serial communication between a compu

Pagina 9

CS650 and CS655 Water Content Reflectometers 6.1 Electrical Specifications Sensor Output: SDI-12 Serial RS-232 Warmup Time: 3 s Measurement Time: 3 m

Pagina 10

CS650 and CS655 Water Content Reflectometers FIGURE 6-1. CS650 and CS655 average current drain FIGURE 6-1 shows average current drain for differen

Pagina 11 - Reflectometers

CS650 and CS655 Water Content Reflectometers Volumetric Water Content using Topp Equation (m3/m3) Range 5% to 50% Accuracy† ±3% VWC typical in min

Pagina 13

CS650 and CS655 Water Content Reflectometers The sensitive volume depends on the surrounding media. In soil, the sensitive volume extends approximat

Pagina 14 - 5. Overview

CS650 and CS655 Water Content Reflectometers 7.3 Datalogger Wiring Our dataloggers typically use SDI-12 to measure the sensor because RS-232 communic

Pagina 15

CS650 and CS655 Water Content Reflectometers and programming examples for Edlog dataloggers are provided at www.campbellsci.com\old-manuals. The SDI1

Pagina 16 - 6. Specifications

CS650 and CS655 Water Content Reflectometers FIGURE 8-1. A200 Sensor-to-PC Interface A data cable, pn 17648, ships with the A200. This cable has a

Pagina 17 - 6.1.1 Current

CS650 and CS655 Water Content Reflectometers 8.1.2 Device Configuration Utility (DevConfig) DevConfig may be downloaded from the Campbell Scientific

Pagina 18

CS650 and CS655 Water Content Reflectometers 8.1.2.1 Settings Editor Tab The Settings Editor tab shows settings stored in the CS650 firmware. Settin

Pagina 19 - 7. Installation

CS650 and CS655 Water Content Reflectometers TABLE 8-2. Real-Time Measurements Measurement Field Name Meaning VWC Volumetric Water Content EC (dS

Pagina 20 - 7.2 Proper Insertion

CS650 and CS655 Water Content Reflectometers 8.1.2.2 Send OS Tab The Send OS tab is used to update the firmware in the CS650. The firmware is availa

Pagina 21 - 7.3 Datalogger Wiring

CS650 and CS655 Water Content Reflectometers TABLE 8-3. CS650 Terminal Commands Command Values Returned Units 0 1) Volumetric Water Content, θ 2)

Pagina 22 - 8. Operation

CS650 and CS655 Water Content Reflectometers 8.2 SDI-12 Measurements The CS650 responds to SDI-12 commands M!, M1!, M2!, M3!, ?!, and I!. TABLE 8-4

Pagina 23

Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service

Pagina 24 - (p. 13))

CS650 and CS655 Water Content Reflectometers 8.2.1 Use of Multiplexers Multiplexers such as Campbell Scientific’s AM16/32B may be used to connect up

Pagina 25

CS650 and CS655 Water Content Reflectometers using an electrical conductivity measurement. A calibration equation converts period and electrical cond

Pagina 26

CS650 and CS655 Water Content Reflectometers between soil solution electrical conductivity and soil bulk electrical conductivity. Most expressions o

Pagina 27

CS650 and CS655 Water Content Reflectometers The effect of temperature on probe electronics is minimal with period average readings varying by less th

Pagina 28

CS650 and CS655 Water Content Reflectometers Quadratic form: θv(Ka) = C0 + C1*Ka + C2*Ka2 with θv the volumetric water content, Ka the bulk dielectri

Pagina 29 - 8.2 SDI-12 Measurements

CS650 and CS655 Water Content Reflectometers derive a calibration depends on the form of the calibration equation. At least three data sets should be

Pagina 30 - Volumetric Water Content

CS650 and CS655 Water Content Reflectometers equilibration is achieved. Collect a set of calibration data values and repeat the water addition proce

Pagina 31 - 8.3.2 The Topp Equation

CS650 and CS655 Water Content Reflectometers • CS650 connected to datalogger programmed to measure probe permittivity • Cylindrical sampling devices

Pagina 32

CS650 and CS655 Water Content Reflectometers The sample tubes should be pushed evenly into the soil surface. Remove the tube and sample and gently t

Pagina 33

CS650 and CS655 Water Content Reflectometers To obtain mwet, subtract the cylinder empty mass and the container empty mass from the mass of the soil f

Pagina 34 - (p. 23)

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Pagina 35

CS650 and CS655 Water Content Reflectometers 9. Maintenance and Troubleshooting The CS650 does not require periodic maintenance. TABLE 9-1 provides

Pagina 36

CS650 and CS655 Water Content Reflectometers 10. References Ledieu, J., P. De Ridder, P. De Clerck, and S. Dautrebande. 1986. “A method of measuring

Pagina 37

CS650 and CS655 Water Content Reflectometers 32

Pagina 38 - 8.4.5 Calculations

Appendix A. Importing Short Cut Code into a Program Editor This tutorial shows: • How to import a Short Cut program into a program editor for addit

Pagina 39

Appendix A. Importing Short Cut Code After pasting, edit the information such that a ' character (single quotation mark) begins each line. This

Pagina 40

Appendix B. Example Programs B.1 CR1000 Programs B.1.1 CR1000 with One CS650 Probe This CRBasic example program measures one CS650 probe on a CR10

Pagina 41 - 10. References

Appendix B. Example Programs Public CS650(6) Public CS650_2(6) DataTable (DatoutCS650,1,-1) DataInterval (0,60,Min,2) Average (3,CS650(1),FP2,Fal

Pagina 42

Appendix B. Example Programs Dim LCount Public CS650(12,6) DataTable (DatoutCS650,1,-1) DataInterval (0,60,Min,2) Average (72,CS650(),IEEE4,False)

Pagina 43

Appendix B. Example Programs 'CS650 #1 SDI12Recorder (CS650(1),"0M3!",1,0) 'CS650 #2 SDI12Recorder (CS650(7),&q

Pagina 44

Appendix C. Discussion of Soil Water Content The water content reflectometer measures volumetric water content. Soil water content is expressed on

Pagina 45 - B.1 CR1000 Programs

Precautions DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO

Pagina 46

Discussion of Soil Water Content C-2

Pagina 47

Appendix D. SDI-12 Sensor Support D.1 SDI-12 Command Basics SDI-12 commands have three components: Sensor address (a) – a single character, and is

Pagina 48

Appendix D. SDI-12 Sensor Support command is 0A2!. The sensor responds with the new address b, which in this case is 2. Send Identification Command

Pagina 49 - Content

Appendix D. SDI-12 Sensor Support Send Data Command (aDv!) This command requests data from the sensor. It is normally issued automatically by the da

Pagina 50

Appendix D. SDI-12 Sensor Support D.2.2 CR200(X) Series Datalogger Example 1. Connect a single CS650 to the datalogger as follows: • Green to Cont

Pagina 51 - D.1 SDI-12 Command Basics

Appendix D. SDI-12 Sensor Support D.2.3 CR1000 Datalogger Example 1. Connect a single CS650 to the datalogger as follows: • Green to Control Port

Pagina 52

Appendix D. SDI-12 Sensor Support D-6

Pagina 54

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe

Pagina 56

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Pagina 57

Table of Contents 8.3.4.2 Insertion Error ..... 22 8.3.5 Temperature Dependence and Corr

Pagina 58

Table of Contents Tables 6-1. Size Specifications .................... 6 6-2. Relative Di

Commenti su questo manuale

Nessun commento