Section 7. Measurement Instructions
Parameter
& Data Type
Enter
SettlingTime
Constant
The time in microseconds to delay between setting up a measurement (switching to the
channel, setting the excitation) and making the measurement. (1 microsecond resolution)
Entry Voltage Range Integration Settling Time
0 All 250 µS 450 µS (default)
0 All _50Hz, _60 Hz 3 mS (default)
>=100 All All µS entered
Integ The time spent on integration in microseconds for each of the channels measured.
Constant Entry
Integration
250
250 µS
_60Hz or 16667
16,667 µS (reject 60 Hz noise)
_50 Hz or 20000
20,000 µS (reject 50 Hz noise)
Mult, Offset
Constant,
Variable,
Array, or
Expression
A multiplier and offset by which to scale the raw results of the measurement.
7.4 Full Bridges
BrFull (Dest, Reps, Range, DiffChan, ExChan, MeasPEx, ExmV, RevEx,
RevDiff, SettlingTime, Integ, Mult, Offset)
X = result w/mult = 1, offset = 0
X
V
V
R
RR
R
RR
x
==
+
+
1000 1000
13
34
2
12
This Instruction applies an excitation voltage to a full bridge and makes a
differential voltage measurement of the bridge output. The resulting value
(multiplier = 1, offset = 0) is the measured voltage in millivolts divided by the
excitation voltage in volts (i.e., millivolts per volt).