CS100 Barometric Pressure Sensor
6. Specifications
Features:
Integral switching circuit limits power consumption to measurement
cycle
Compatible with all Campbell Scientific dataloggers (including the
CR200(X) series)
Calibration NIST traceable
Meets CE conformance standards
Compatibility
Dataloggers: CR200(X) series
CR800 series
CR1000
CR3000
CR5000
CR9000(X)
CR7X
CR510
CR10(X)
CR23X
21X
6.1 Performance
6.1.1 Performance for “Standard” Range Option
Measurement Range: 600 mb to 1100 mb (hPa)
Operating Temperature Range: -40° to +60°C (-40° to +140°F)
Storage Temperature Range: -60° to +120°C (-76° to +248°F)
Proof Pressure: 1500 mb
Burst Pressure: 2000 mb
Humidity Range: non-condensing (up to 95% RH)
Media Compatibility: non-corrosive, non-condensing air or gas
Resolution: 0.01 mb
Total Accuracy
1
: ±0.5 mb @ 20°C
±1.0 mb @ 0° to +40°C
±1.5 mb @ -20° to +50°C
±2.0 mb @ -40° to +60°C
Linearity: ±0.4 mb
Hysteresis: ±0.05 mb
Repeatability: ±0.03 mb
Long-term Stability: ±0.1 mb per year
1
The root sum squared (RSS) of end point non-linearity, hysteresis, non-
repeatability and calibration uncertainty.